The specified measurement range is 21% to 100% oxygen (O2). Values outside this range are outside the declared measurement envelope and may not be reliably reported.
The unit uses an ENVITEC OOM102-1 galvanic fuel cell sensor. The sensor life is specified as 1,000,000 O2-percent hours (typical service life in normal use).
Typical response time is about 15 seconds to reach a reading representative of the current oxygen concentration.
The analyzer provides configurable high and low alarms with both visual and audio (buzzer) indications. The device requires a minimum 1% separation (hysteresis) between high and low setpoints to prevent rapid toggling; the high alarm triggers when the measured value exceeds the configured high setpoint and the low alarm triggers when the measured value falls below the configured low setpoint.
Yes. The analyzer is compact and lightweight: approximately 84 mm x 130 mm x 31 mm and about 0.25 kg, making it convenient for handheld or portable use.
The device is powered by three AAA alkaline batteries. Battery status is displayed on the LCD screen with a low battery alert to indicate when batteries should be replaced.
The analyzer has an IPX2 rating, which provides protection against vertically dripping water when the device is tilted up to 15 degrees. It is not fully water‑proof and should not be exposed to heavy moisture or submerged.
Yes. The product is suitable for healthcare settings (e.g., monitoring oxygen supply), industrial environments where oxygen concentration must be controlled, and environmental or research applications. Always ensure the device meets any regulatory or clinical requirements specific to your application before relying on it for critical decisions.
The galvanic sensor requires minimal maintenance, but periodic checks and calibration are recommended to ensure accuracy. Specific calibration procedures and intervals are provided in the product manual; contact your supplier or service center for calibration support and replacement sensors.
The LCD shows the real-time oxygen concentration reading, alarm indicators (high/low), battery status/low battery alert, and any alarm or fault notifications to facilitate easy monitoring.
Yes. High and low alarm conditions are supported with an audible buzzer in addition to on-screen alarm indications.
The product features a precise alarm system and a reliable galvanic sensor for accurate readings in its specified range. For specific numerical accuracy and uncertainty figures, refer to the technical datasheet or contact the manufacturer/supplier for detailed performance specifications.
Contact your authorized dealer or the manufacturer for replacement sensors (ENVITEC OOM102-1) and repair/service information. Do not attempt to modify or repair the sensor electronics yourself; follow the service instructions in the manual.