The BDF-40H105 offers 105 liters of usable storage capacity, designed in a horizontal layout to maximize space and ease of access for lab samples and medical products.
The unit supports temperatures down to -40°C. The model's operating range is specified as approximately -10°C to -40°C — exact achievable setpoints may depend on ambient conditions and load.
Temperature is managed by a microprocessor control system and shown on a clear LED display that lets you monitor and adjust setpoints with 0.1°C accuracy.
Yes. The system includes built-in alarms for abnormal conditions such as high/low temperature, sensor faults, power failure and controller faults. Please refer to the user manual for the specific alarm behavior for your unit.
This model uses a direct refrigeration system with R290 refrigerant and a copper evaporator for efficient heat transfer.
External dimensions are 680 × 580 × 935 mm (L × W × H). Gross weight (packaged) is approximately 65 kg; packaged size is about 750 × 700 × 1040 mm.
The interior is made of 304 stainless steel for corrosion resistance and easy cleaning. The exterior is cold-rolled steel coated with an anti-bacterial powder finish. Insulation is rigid polyurethane foam in a unibody construction.
Yes. The unit includes a safety lock to restrict unauthorized access, a durable hygienic interior, and alarm features suitable for medical and laboratory storage requirements.
Power supply is AC 220V ±10%, 50/60Hz (110V, 60Hz option available). Typical power consumption for the 105L model is approximately 160 W.
The BDF-40H105 is designed for quiet operation appropriate for lab and clinical settings. Exact decibel ratings are not specified; if you need precise noise data, request the technical sheet from the supplier.
Regular maintenance includes keeping the condenser and vents clean, cleaning the interior with appropriate disinfectant, checking door seals, verifying temperature accuracy with an independent thermometer, and servicing the refrigeration system per the manufacturer's schedule. Defrosting approach (manual or automatic) is model-dependent — consult the manual.
Yes. The freezer is engineered to maintain low, stable temperatures required for many vaccines and biological samples. Always validate storage conditions and follow specific vaccine or sample manufacturer recommendations and regulatory guidelines.
The listed features include a microprocessor controller with LED display but do not specify built-in data logging or remote monitoring. External data loggers or optional monitoring systems can typically be integrated — check with the supplier for compatible accessories.
Most units come with standard internal shelving or baskets; optional racks and customized inserts may be available. Confirm the included accessories and ordering options with your vendor.
Install on a level, stable surface with adequate clearance for ventilation (follow the manual for recommended clearances). Keep away from heat sources and ensure correct voltage supply. Because R290 is a flammable refrigerant, installation and service should be performed by qualified technicians in accordance with local codes.